Categories
LGMI-16
LGMI-16 intrinsically safe teletechnical meter
LGMI-16 intrinsically safe teletechnical meter - a universal measuring device used to test the parameters of underground telecommunication networks and other electrical parameters.
The LGMI-16 intrinsically safe teletechnical meter is a universal measuring device used to test underground parameters of telecommunication networks and other electrical parameters. It enables, among others, assessment of the efficiency of telecommunication networks and the size of interference occurring in the tracks. The device allows for measurements of such parameters as: Alternating and direct voltage Alternating and direct current Resistance Signal attenuation The meter also allows for generation of a sinusoidal signal with a given amplitude and frequency. Measurements are made using the "True RMS" method
- DC voltmeter
Range | Resolution | Accuracy |
200mV | 100μV | ±0.5% ±1 digit |
2V | 1mv | ±0.5% ±1 digit |
20V | 10mV | ±0.5% ±1 digit |
60V | 100mV | ±0.5% ±1 digit |
- AC voltmeter (50 – 3400 Hz)
Range | Resolution | Accuracy |
200mV | 100μV | ±1.2% ±3 digits |
2V | 1mv | ±0.8% ±3 digits |
20V | 10mV | ±0.8% ±3 digits |
60V | 100mV | ±0.8% ±3 digits |
- Level meter (300 – 3400 Hz)
- From 2mV (-51.76 dB) to 200mV (-11.76 dB)
- From 200mV (-11.76 dB) to 2V (8.24 dB)
- 0dB = 775mV
- DC Ammeter
Range | Resolution | Accuracy |
2A | 1mA | ±0.5% ±3 digits |
Input impedance: max. 1Ω. Possibility of installing an external shunt
- DC milliammeter
Range | Resolution | Accuracy |
2mA | 1μA | ±0.8% ±3 digits |
20mA | 10uA | ±0.8% ±3 digits |
200mA | 100uA | ±0.5% ±3 digits |
Input impedance: max. 1Ω. Possibility of installing an external shunt
- AC ammeter (50 – 3400 Hz)
Range | Resolution | Accuracy |
2A | 1mA | ± 3% ± 3 digits |
- AC milliammeter (50 – 3400 Hz)
Range | Resolution | Accuracy |
2mA | 1μA | ± 1% ± 3 digits |
20mA | 10uA | ± 1% ± 3 digits |
200mA | 100uA | ±2% ±3 digits |
- Ohmmeter
Range | Resolution | Accuracy |
200Ω | 0.1Ω | ±0.8% ±3 digits |
2kΩ | 1Ω | ±0.8% ±1 digit |
20kΩ | 10 Ohm | ±0.8% ±1 digit |
200kΩ | 100Ω | ±0.8% ±1 digit |
2MΩ | 1kΩ | ±0.8% ±1 digit |
- Gigaohmmeter
Range | Resolution | Accuracy |
20MΩ | 10kΩ | ± 2% ± 2 digits |
200MΩ | 100kΩ | ± 2% ± 2 digits |
2GΩ | 1MΩ | ± 10% ± 2 digits |
- Generator
- Frequency range 300 - 3400Hz
- Voltage range: 1mV - 1V
- Circuit Continuity Test (with sound)
ADDITIONAL FEATURES
- Large graphic display showing precise measurement parameters
- Intuitive menu
- Handy, compact design